Alarm thresholds set an acceptable range of values for a statistic. Statistics that exceed the normal threshold raise an alarm.
To set alarm thresholds
Click Configure | Alarms.
Set the range of severities you want to apply to the alarm. Use the threshold bar or the severity table.
Threshold bar
|
Move the mouse pointer over the threshold between two severities until it appears as a split-bar pointer. Click and drag the mouse to move the threshold to a new position. When you have completed the move, the change also appears in the severity table. |
Severity table
|
Click in the Start cell for each chosen severity. Type the lower value for a required threshold range. Ensure the start values for each range are consistent. A failure to do so may cause problems when alarms are raised. |
Note: Not all alarms have thresholds.